First In-Human Experience using a Novel Hybrid Device to simplify Left Subclavian Artery Anastomosis during Total Arch Replacement

Presented During:

Thursday, November 19, 2026: 6:03AM - Friday, November 20, 2026: 6:03PM
Sheraton Times Square  
Posted Room Name: Exhibit Hall & Metropolitan Ballroom  

Abstract No:

17430 

Submission Type:

Abstract Submission 

Authors:

Leonard Pitts (1), Matteo Montagner (1), Markus Kofler (1), Julius Kaemmel (1), Gaik Nersesian (1), Elisa Jankwitz (1), Semih Buz (1), Volkmar Falk, MD (1), Jörg Kempfert (1), Malakh Shrestha (2)

Institutions:

(1) Deutsches Herzzentrum der Charité, Berlin, Germany, (2) Mayo Clinic, Rochester, MN

Submitting Author:

Leonard Pitts    -  Contact Me
Deutsches Herzzentrum der Charité

Co-Author(s):

Matteo Montagner    -  Contact Me
Deutsches Herzzentrum der Charité
Markus Kofler    -  Contact Me
Deutsches Herzzentrum der Charité
Julius Kaemmel    -  Contact Me
Deutsches Herzzentrum der Charité
Gaik Nersesian    -  Contact Me
Deutsches Herzzentrum der Charité
Elisa Jankwitz    -  Contact Me
Deutsches Herzzentrum der Charité
Semih Buz    -  Contact Me
Deutsches Herzzentrum der Charité
*Volkmar Falk, MD    -  Contact Me
Deutsches Herzzentrum der Charité
Jörg Kempfert    -  Contact Me
Deutsches Herzzentrum der Charité
*Malakh Shrestha    -  Contact Me
Mayo Clinic

Presenting Author:

Leonard Pitts    -  Contact Me
Deutsches Herzzentrum der Charité

Abstract:

Objectives: Total arch replacement using the Frozen Elephant Trunk (FET) technique remains the gold standard for a definite aortic arch repair. The supra-aortic vessel anastomoses, especially those to the left subclavian artery (LSA), can be technically challenging, resulting in the need for longer circulatory arrest and increased perioperative risk. We report the first experience and outcome in two patients treated with a new hybrid device to simplify the anastomosis of the LSA during total aortic arch replacement.

Methods: Two female patients (Case 1: 69 years old, combined ascending (56mm) and descending (50mm) aortic aneurysm, Euroscore II 9.28%; Case 2: 74 years old, penetrating aortic ulcers of the distal aortic arch, status post mechanical Bentall root replacement, Euroscore II 13.87%) underwent elective total aortic arch replacement using the FET technique in Octobre 2025. Distal anastomosis was performed in zone 2 during circulatory arrest at 28°C and bilateral antegrade cerebral perfusion. The novel hybrid device RapidLink – consisting of a distal stent portion connected to a proximal dacron graft – was inserted into the transected LSA under direct vision. Sizing of the device was based on the LSA diameter and length from the LSA ostium to the left vertebral artery ostium.

Results: No device-related complications were observed. Both patients experienced an uneventful ICU stay and were discharged home on the 6th (Case 1: aspirin + clopidogrel) and 14th (Case 2: marcumar) postoperative day with individualized oral anticoagulation. Computed tomography angiography (CTA) prior discharge showed satisfactory results and no signs of endoleak. During follow-up, CTA revealed no signs of stent thrombosis (Case 1: CTA and planned TEVAR extension after 3 months, Case 2: CTA after 5 months).

Conclusions: Anastomosis of the LSA using the novel hybrid device RapidLink during total arch replacement is safe and feasible. During follow-up, CTA shows satisfactory results with no signs of endoleak or thrombus. The device has great potential to reduce complexity of total arch replacement by simplifying the LSA anastomosis.

Aortic Symposium:

Aortic Arch

 

Keywords - Adult

Aorta - Aorta
Aorta - Aortic Arch