Development of real-time, high-fidelity physiologic cerebral monitoring during aortic arch surgery

Presented During:

Thursday, November 19, 2026: 6:03AM - Friday, November 20, 2026: 6:03PM
Sheraton Times Square  
Posted Room Name: Exhibit Hall & Metropolitan Ballroom  

Abstract No:

17541 

Submission Type:

Abstract Submission 

Authors:

Lauren Levy (1), Srdjan Dragovic (1), Jason Glotzbach (1), Christian Guay (1), Andrea Steely (1)

Institutions:

(1) University of Utah, Salt Lake City, UT

Submitting Author:

Lauren Levy    -  Contact Me
University of Utah

Co-Author(s):

Srdjan Dragovic    -  Contact Me
University of Utah
Jason Glotzbach    -  Contact Me
University of Utah
Christian Guay    -  Contact Me
University of Utah
Andrea Steely    -  Contact Me
University of Utah

Presenting Author:

Lauren Levy    -  Contact Me
University of Utah

Abstract:

Objective: While Near-Infrared Spectroscopy (NIRS) cerebral oximetry is the most common method for estimating regional cerebral perfusion during circulatory arrest (CA) for aortic surgery, it is a low-fidelity technique that may not fully capture frontal brain cerebral activity. Frontal electroencephalogram (EEG), however, directly monitors frontal cortex function. We assessed frontal EEG compared to NIRS cerebral oximetry during elective aortic arch surgery as a high-fidelity method for monitoring cerebral perfusion.
Methods: Five patients underwent elective ascending aorta plus total or partial arch replacement at a single institution from April-July 2026. Cardiopulmonary bypass (CPB) was established via aorta-right atrial cannulation; CA was performed at moderate hypothermia (24-28°C) with bilateral antegrade cerebral perfusion (ACP). Frontal EEG and NIRS cerebral oxygen saturation (StO2) were monitored simultaneously throughout surgery.
Results: Four patients underwent hemiarch and one underwent total arch replacement. Mean age was 56.4 ± 14.5 years. 60% were male. Mean BMI was 30.9 ± 7.4 kg/m2. 80% had hypertension, hyperlipidemia, and were nonsmokers. No patients had cerebrovascular disease or carotid stenosis. Two patients underwent bio-Bentall root replacement for moderate+ aortic stenosis (AS); one had an aortic valve replacement for mixed moderate AS/aortic insufficiency (AI), and one underwent aortic valve repair for mild AI. The total arch patient also underwent frozen elephant trunk placement. Mean CPB, aortic cross-clamp, and CA times were 246.4 ± 26.0, 142.2 ± 50.1, and 26.0 ± 10.1 minutes, respectively. During cooling and CA with ACP, frontal EEG demonstrated a change in brain activity from broad spectral activity to predominantly low-frequency activity corresponding to delta and theta waves in four patients, indicating slowing of frontal cortical electrical activity and correlating with observed decreases in NIRS cerebral StO2. In Patient #3, frontal EEG demonstrated burst suppression correlating to intra-operative stroke that was not detected by NIRS.
Conclusions: Frontal EEG may provide more sensitive, real-time cerebral monitoring than NIRS cerebral oximetry during aortic arch surgery including capture of massive neurological events. This pilot study demonstrates equipoise for using frontal EEG as a standardized cerebral monitoring method in future prospective trials investigating neurologic outcomes during aortic arch surgery.

Aortic Symposium:

Anesthesia and Perioperative Management

 

Keywords - Adult

Aorta - Aorta
Aorta - Aortic Arch
Perioperative Management/Critical Care - Perioperative Management