Reperfusion Without Revascularization: Resolution of Lower Extremity Malperfusion Following Type A Aortic Dissection Repair With Ascyrus Medical Dissection Stent
Presented During:
Thursday, November 19, 2026: 6:03AM - Friday, November 20, 2026: 6:03PM
Sheraton Times Square
Posted Room Name:
Exhibit Hall & Metropolitan Ballroom
Abstract No:
17608
Submission Type:
Abstract Submission
Authors:
Lauryn Ridley (1), Anna Marie Harding (1), Adanna Akujuo (2)
Institutions:
(1) University of Virginia Medical School, Charlottesville, VA, (2) University of Virginia Health System University Hospital, Charlottesville, VA
Submitting Author:
Lauryn Ridley
-
Contact Me
University of Virginia Medical School
Co-Author(s):
Anna Marie Harding
-
Contact Me
University of Virginia Medical School
Adanna Akujuo
-
Contact Me
University of Virginia Health System University Hospital
Presenting Author:
Lauryn Ridley
-
Contact Me
University of Virginia Health System University Hospital
Abstract:
Objective:
Malperfusion complicating acute Type A aortic dissection may persist after proximal repair. We report resolution of severe bilateral lower-extremity malperfusion using the Ascyrus Medical Dissection Stent (AMDS) without additional revascularization.
Methods:
A 74-year-old woman presented with acute back pain. CTA demonstrated a Stanford Type A dissection extending into the descending aorta, absent flow in the left femoral artery, severe narrowing of the right femoral artery, and an ascending aortic intimal tear with contained rupture. Examination revealed absent bilateral femoral Doppler signals, mild mottling, and complete loss of motor and sensory function in the left leg. Preoperative CTA measurements supported use of a 40-mm AMDS. Emergent repair was performed with ascending aortic and venous cannulation, cooling to 18°C, retrograde cerebral perfusion, and circulatory arrest. A contained ascending aortic rupture tamponaded by the superior vena cava was identified. The distal aorta was resected to a 2-cm cuff proximal to the innominate artery; arch inspection showed no additional tear. The AMDS was deployed under direct visualization without compromising the innominate artery and secured with an external felt strip and running 3-0 polypropylene suture. A 32-mm Gelweave graft with a side arm was anastomosed end-to-end to the aorta and AMDS. Cardiopulmonary bypass was resumed through the side arm, and proximal graft anastomosis was completed above the sinotubular junction.
Results:
Bilateral femoral pulses returned immediately, with complete resolution of mottling. Both calves remained soft; neither fasciotomy nor open or endovascular limb revascularization was required. Postoperative CTA confirmed false-lumen thrombosis and substantial true-lumen expansion in both common iliac arteries, with preserved distal runoff. Residual false-lumen filling was confined to the right common carotid artery.
Conclusions:
In this patient with limb-threatening malperfusion, AMDS implantation during central Type A dissection repair produced immediate reperfusion without adjunctive revascularization. By stabilizing the distal dissected aorta, expanding the true lumen, and promoting false-lumen thrombosis, AMDS addresses the dynamic mechanism of malperfusion. Its rapid, reproducible deployment may reduce additional procedures while limiting circulatory-arrest and ischemic time in this critically ill population.
Aortic Symposium:
Dissection
Keywords - Adult
Adult
Aorta - Aortic Disection
Perioperative Management/Critical Care - Perioperative Management
You have unsaved changes.