Multimodal Auscultation and ECG Foundation Model Enables Point-of-Care Bicuspid Aortic Valve Detection

Presented During:

Thursday, November 19, 2026: 6:03AM - Friday, November 20, 2026: 6:03PM
Sheraton Times Square  
Posted Room Name: Exhibit Hall & Metropolitan Ballroom  

Abstract No:

17695 

Submission Type:

Abstract Submission 

Authors:

Alokkumar Jha (1), Jennifer Kwan (2), Colie Sutter (3), George Mathew (3)

Institutions:

(1) Weill Cornell Medical College, Cornell University, New York, NY, (2) Yale University School of Medicine, New Haven, CT, (3) Eko Health Inc., Emeryville, CA

Submitting Author:

Alokkumar Jha    -  Contact Me
Weill Cornell Medical College, Cornell University

Co-Author(s):

Jennifer Kwan    -  Contact Me
Yale University School of Medicine
Colie Sutter    -  Contact Me
Eko Health Inc.
George Mathew    -  Contact Me
Eko Health Inc.

Presenting Author:

Alokkumar Jha    -  Contact Me
Weill Cornell Medical College, Cornell University

Abstract:

Objective
Develop and evaluate a multimodal foundation model ensemble for bicuspid aortic valve detection using simultaneous cardiac auscultation and electrocardiogram from a point-of-care digital stethoscope.

Methods
Retrospective case-control study of 124 matched patients (62 bicuspid aortic valve, 62 controls) with simultaneous cardiac recordings across nine positions. Dataset: 1,047 recordings, 13,627 beats. Pipeline: (1) beat segmentation; (2) Mel Frequency Cepstral Coefficient extraction; (3) self-supervised pretraining via masked modeling and contrastive learning; (4) foundation model embedding aggregation; (5) ensemble combining Mel Frequency Cepstral Coefficient-based and foundation model support vector machines. Performance evaluated by 5-fold stratified cross-validation with 2,000 bootstrap resamples.

Results
Multimodal ensemble achieved area under curve 0.756 (95% confidence interval 0.665–0.839), sensitivity 67.7% (42 of 62), specificity 72.6% (45 of 62), positive predictive value 71.2%, negative predictive value 69.2%. Mel Frequency Cepstral Coefficient-only classifier: area under curve 0.712. Label-free foundation model alone: area under curve 0.677, demonstrating self-supervised pretraining captured clinically meaningful cardiac structure without disease labels. Ensemble outperformed either modality individually. Feature analysis identified beat-to-beat spectral variability at mitral and tricuspid positions as most discriminative (Figure 1). Power spectral density analysis revealed significantly elevated acoustic energy in 50–400 Hertz murmur band across 112 bicuspid aortic valve recordings at aortic notch compared to 122 controls (Figure 1).

Conclusions
Point-of-care simultaneous cardiac auscultation and electrocardiogram contain sufficient multimodal information for clinically meaningful bicuspid aortic valve screening. Self-supervised foundation models achieved near-equivalent performance to engineered features, validating foundation model pretraining captures clinically relevant structure without disease-specific supervision. Ensemble superiority confirms complementary physiological information in each modality. Beat-to-beat spectral variability emerged as quantitative acoustic signature of murmur radiation. These findings support deployment of multimodal cardiac artificial intelligence screening in primary care to triage patients toward confirmatory echocardiography.

Aortic Symposium:

Aortic Root

Image or Table

Supporting Image: BAV_Figure1_AcousticPhenotype.png

Presentation

BAV_AATS_Presentation.pptx
 

Keywords - Adult

Aorta - Aortic Root
Aorta - Ascending Aorta