Abstract:
Objective: Surveillance of chronic type B aortic dissection (TBAD) requires assessment of aortic diameter and false lumen patency. Lumen-specific hemodynamics may be relevant to aortic remodelling and late risk of rupture, or organ malperfusion. Ultrasound technology, incorporating novel vector flow imaging (V Flow), facilitates true (TL) and false (FL) lumen haemodynamic assessment of medically managed TBAD. We aimed to assess utility of V Flow imaging in sequential TBAD ultrasound studies
Methods: Four patients with conservatively managed chronic TBAD were prospectively studied with sequential transabdominal ultrasound from September 2025 to July 2026. B-mode and Doppler ultrasound measurements of abdominal aortic diameter, flap mobility, FL thrombus, and TL/FL peak systolic velocities and flow direction were performed. Adjunct V Flow was feasible in two lower- BMI patients, providing TL/FL volume flow, TL/FL wall shear stress (WSS), and tear-site turbulence index. Comparison CT angiography or 4D flow MRI hemodynamic parameters were available for three patients.
Results: In patient 1, V Flow evidenced FL-dominant infrarenal perfusion at baseline (TL:FL 0.62) becoming TL-dominant at 9 months (TL:FL 4.4) despite stable diameter, with increasing TL and decreasing FL wall shear stress. Patient 2 had persistently TL-dominant flow and high entry-tear turbulence, consistent with CT findings of a narrow tear and partial FL thrombosis. In two higher-BMI patients (3 and 4), V Flow was not feasible, but transabdominal ultrasound demonstrated TL expansion with progression to complete FL thrombosis in one and stable diameters with persistent partial FL thrombus in another. CT/4D flow MRI corroborated the hemodynamic patterns in patients 3 and 4.
Conclusions: Serial transabdominal ultrasound, with adjunct V Flow when feasible, provides bedside TL/FL hemodynamic assessment in chronic TBAD, complementing CT and 4D flow MRI. This series demonstrated that abdominal flow remodelling from FL‑ to TL‑dominant perfusion could be detected, even when aortic diameters remained stable. As a noninvasive imaging modality, not requiring intravenous contrast agent, V Flow ultrasound may augment the utility of ultrasound surveillance of medically managed TBAD patients.