Adapting to the AMDS aortic prosthesis, lessons learned from re-operative interventions

Presented During:

Thursday, November 19, 2026: 6:03AM - Friday, November 20, 2026: 6:03PM
Sheraton Times Square  
Posted Room Name: Exhibit Hall & Metropolitan Ballroom  

Abstract No:

17714 

Submission Type:

Abstract Submission 

Authors:

Evangelos Anastasakis (1), Firas Aljanadi (2), Rickesh Karsan (2), Susan Durkan (2), Mark Boyle (2), Pushpinder Sidhu (2), Christopher Austin (2), Alsir Ahmed (2), Reuben Jeganathan (2), Katie O' Sullivan (2)

Institutions:

(1) St George’s University Hospital NHS Foundation Trust London, London, United Kingdom, (2) Royal Victoria Hospital, Belfast, United Kingdom

Submitting Author:

Evangelos Anastasakis    -  Contact Me
St George’s University Hospital NHS Foundation Trust London

Co-Author(s):

Firas Aljanadi    -  Contact Me
Royal Victoria Hospital
Rickesh Karsan    -  Contact Me
Royal Victoria Hospital
Susan Durkan    -  Contact Me
Royal Victoria Hospital
Mark Boyle    -  Contact Me
Royal Victoria Hospital
Pushpinder Sidhu    -  Contact Me
Royal Victoria Hospital
Christopher Austin    -  Contact Me
Royal Victoria Hospital
Alsir Ahmed    -  Contact Me
Royal Victoria Hospital
Reuben Jeganathan    -  Contact Me
Royal Victoria Hospital
Katie O' Sullivan    -  Contact Me
Royal Victoria Hospital

Presenting Author:

Evangelos Anastasakis    -  Contact Me
St George’s University Hospital NHS Foundation Trust London

Abstract:

Objective:
The Ascyrus Medical Dissection Stent (AMDS) is increasingly used as an adjunct in emergency repair of acute type A aortic dissection (ATAAD). It aims to reduce the incidence of distal anastomotic new entry (DANE) tears and false-lumen expansion, with favorable 5-year outcomes. However, the device is not easily explanted and evidence guiding re-intervention is limited.

Methods:
A retrospective analysis of consecutive patients undergoing AMDS implantation for emergency ATAAD repair at a single centre since November 2021 was performed. Outcomes were 30-day mortality, ICU and hospital length of stay, re-sternotomy for bleeding, new stroke and re-intervention. In patients requiring re-intervention, indication, interval from index surgery, operative strategy, management of the existing AMDS, peri-operative complications and outcomes at follow-up were reviewed.

Results:
Thirteen patients underwent AMDS implantation. There was no 30-day mortality. Median ICU and hospital stay were 7 (5–9) and 21 (13–23) days, respectively. Re-sternotomy for bleeding occurred in 1 (7.7%) and new stroke in 2 (15.4%). Three (23.1%) required re-intervention. Case 1 developed transfusion-dependent haemolysis, renal impairment and false-lumen expansion. At 9-month redo surgery, the proximal AMDS cuff was stenosed to ~1 cm; the AMDS was cut back to Zone 1, permitting total arch replacement with frozen elephant trunk (FET). Haemolysis and renal impairment resolved. Case 2 developed graft infection 7 months after ATAAD repair. Management comprised extensive debridement and serial mediastinal washouts, with ultimate omental flap coverage. There was good recovery on long-term suppressive antibiotics. Case 3 developed persistent false-lumen perfusion with a peri-graft collection enlarging from 4.5 to 6.5 cm over 8 months. Redo total arch replacement was performed with FET deployment through the existing AMDS. The descending thoracic aorta subsequently enlarged from 3.7 to 6.5 cm over 14 months and thoracic endovascular aortic repair was performed.

Conclusions:
Re-intervention after AMDS requires tailored strategies. FET deployment through a patent AMDS is feasible but technically challenging; cuff stenosis may require cut-back; and graft infection may be managed with debridement, washout and omental flap coverage. Encouraging 5-year outcomes support wider uptake; careful case selection and familiarity with re-operative strategies are essential to optimise its use.

Aortic Symposium:

Dissection

 

Keywords - Adult

Aorta - Aortic Disection