Objective: Transcatheter aortic valve replacement (TAVR) valve embolization is a rare but known complication that occurs most commonly in the ascending aorta during deployment. Salvage of an embolized balloon-expandable valve often requires distal repositioning and stenting within the aorta, which carries risk of aortic dissection. We present a case of iatrogenic, asymptomatic type B aortic dissection (TBAD) associated with percutaneous repositioning of an embolized TAVR valve.
Methods: This is a case report of an 83-year-old man with severe aortic stenosis and prohibitive surgical risk who underwent TAVR with a 29mm valve. Loss of capture during deployment resulted in valve embolization into the ascending aorta. He remained hemodynamically stable, and the embolized valve was repositioned and balloon-expanded distal to the left subclavian artery. A second valve was then successfully implanted in the native annulus. Post-TAVR computed tomography angiography (CTA) demonstrated stable positioning of the embolized valve at the aortic isthmus but revealed a new adjacent TBAD extending into the thoracoabdominal aorta. The patient remained asymptomatic from dissection and was managed medically with surveillance. One month later, CTA demonstrated false lumen expansion and a new 1.5cm pseudoaneurysm adjacent to the embolized TAVR valve, prompting cardiac surgery evaluation and repair.
Results: Following left carotid-subclavian bypass, zone 1 TEVAR was performed using a branched TEVAR device with the side branch placed into the left subclavian artery. The branched TEVAR body was deployed through/adjacent to the previously embolized TAVR valve and carefully apposed without disrupting it. Completion angiography demonstrated exclusion of the pseudoaneurysm and false lumen flow, patent head vessels, and no endoleak. The patient was extubated in the operating room, left the ICU on postoperative day 2, and was discharged home on day 5. Six-month and one-year CTA demonstrated complete coverage of arch pseudoaneurysm and healing.
Conclusions: Distal fixation of an embolized TAVR valve can cause occult aortic wall injury; careful, gentle apposition may reduce this risk. This case highlights the complex management of TAVR embolization complicated by iatrogenic type B aortic dissection. Given the location of the embolized TAVR valve, branched TEVAR with head vessel debranching is a feasible option with favorable early outcomes. Continued imaging surveillance is essential.
Authors
Bo Chang Wu (1), Branson Taheri (2), Elizabeth Bashian (3), Meng Han Tsai (4), T. Brett Reece (5), Muhammad Aftab (6)
Institutions
(1) University of Colorado Anschutz, Aurora, CO, (2) University of Colorado Anschutz Medical Center, Denver, CO, (3) University of Colorado Anschutz Medical Center, Aurora, CO, (4) University of Colorado, Anschutz Medical Center, Aurora, CO, Aurora, CO, (5) University of Colorado Hospital, Aurora, CO, (6) University of Colorado Anschutz Medical Center and Rocky Mountain Regional VAMC, Aurora, CO
Aortic Symposium
Endoluminal Prostheses