Association Between Thoracic Stent-Graft Design and Major Proximal Device Failure After Hybrid Aortic Arch Repair

Victor Mosquera Rodriguez Poster Presenter
Department of Cardiac Surgery, University Hospital of A Coruña, A Coruña, Spain
A Coruña
Spain
 - Contact Me

Víctor X. Mosquera Rodríguez, MD, PhD is Section Head of Adult Cardiac Surgery and Director of the Heart Transplant Program at the University Hospital A Coruña (CHUAC), Spain. He is also Associate Professor of Surgery at the University of Santiago de Compostela.

His clinical and research interests focus on aortic surgery, structural heart interventions, transcatheter valve therapies, hybrid cardiovascular procedures, and heart transplantation. He has extensive experience in complex aortic surgery and has pioneered one of Europe’s largest programs in non-transfemoral transcatheter aortic valve implantation (TAVI), with particular expertise in transcarotid and alternative-access techniques.

Dr. Mosquera has authored numerous peer-reviewed publications and regularly serves as invited faculty and speaker at major international meetings, including AATS, EACTS, PCR, and other cardiovascular congresses. His current research is centered on hybrid aortic arch repair, endovascular treatment of the ascending aorta, transcatheter electrosurgery, and the optimization of alternative-access structural heart interventions.

He actively collaborates with international research groups and industry partners in the development and evaluation of innovative cardiovascular technologies, while maintaining a strong commitment to surgical education, clinical excellence, and translational research.

Thursday, November 19, 2026: 6:03 AM - Friday, November 20, 2026: 6:03 PM
Sheraton Times Square 
Room: Exhibit Hall & Metropolitan Ballroom 

Description

Objective: To evaluate whether thoracic stent-graft design is associated with major proximal device-related failure after hybrid aortic arch repair combining supra-aortic debranching or translocation with thoracic endovascular aortic repair.
Methods: We retrospectively analyzed 110 consecutive high-risk patients who underwent conventional hybrid aortic arch repair between January 2010 and January 2025. Procedures consisted of supra-aortic debranching or translocation followed by thoracic endovascular aortic repair with proximal landing in Ishimaru zones 0–2. The primary endpoint was major proximal device-related failure, defined as type Ia or III endoleak, retrograde type A dissection, endograft infection, or aortobronchial fistula requiring reintervention. In-hospital mortality was analyzed separately. Device-related event rates were compared using Fisher's exact test, followed by an exploratory Cochran-Armitage trend analysis. Event-free survival was assessed using Kaplan-Meier analysis and the log-rank test.
Results: The cohort included 110 patients (mean age 65.0 years; 88/110 [80%] men) with a mean follow-up of 69.3 months. In-hospital mortality was 6/110 (5.5%) and stroke occurred in 3/110 patients (2.7%). Major proximal device-related failure occurred in 16/110 patients (14.5%), comprising 10 type Ia or III endoleaks, 3 retrograde type A dissections, 2 endograft infections, and 1 aortobronchial fistula; all required reintervention (8 open, 8 endovascular). Event rates were 1/30 (3.3%), 5/34 (14.7%), and 10/46 (21.7%) across the three stent-graft design groups. Compared with the pooled comparator group, the lowest-event design was associated with a lower risk of major proximal device-related failure (odds ratio 0.15; 95% confidence interval 0.02–0.97; P=0.041). An exploratory trend analysis demonstrated increasing event rates across designs (P=0.011). Event-free survival also differed significantly among groups (log-rank P=0.020).
Conclusions: Major proximal device-related failure was the principal determinant of late treatment failure after conventional hybrid aortic arch repair. Thoracic stent-graft design was associated with different rates of proximal device-related failure and reintervention. These hypothesis-generating findings support further multicenter studies to determine whether device design independently influences long-term durability after hybrid aortic arch repair.